EasyManuals Logo
Home>Tektronix>Oscilloscope>TDS2014C

Tektronix TDS2014C User Manual

Tektronix TDS2014C
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #108 background imageLoading...
Page #108 background image
Maintenance
Table 6-10: Expected signals at J202 (cont.)
Signal at J202 Description
10 FPDATA_34_P
1
Front Panel Scan Data line B No xed pattern. Stable when
triggered from FPCLR_34_P. Pattern depends on rotation of
knobs being scanned.
11 FPCLR_34_P
2
Front Panel Clock line B 405 kHz. High with 50 ns pulses going
low. Each clock after FPCLR_34_P increments the scan location
by one.
12 FPCLK_34_P
2
Clear Front Panel Scan line B 6.44 kHz. Low pulsing high with
50 ns pulses. Determines beginning of a scan sequence for
second half of front panel.
1
Unused and open on 2-channel models.
2
Unused and tied low on 2-channel models.
11. If all of the signals are present, the Front-Pane l board is probably defective.
Replace it.
12. If some or all of the signals a re missing, continue with step 13.
13. Turn off the oscilloscope and disconnect the front-panel cable from the Main
board at connector J202.
14. Turn on the oscilloscope and check connector J202 for the same signals as
in step 3.
15. If some or all of the signals are missing, the Main board is probably defective.
Replace it.
6–36 TDS2000C Series Oscilloscope Service Manual

Table of Contents

Other manuals for Tektronix TDS2014C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS2014C and is the answer not in the manual?

Tektronix TDS2014C Specifications

General IconGeneral
BrandTektronix
ModelTDS2014C
CategoryOscilloscope
LanguageEnglish

Related product manuals