Operation 3-3
Warm - up T im e
An initial warm-up period of 15 minutes is recommended before sampling
with the Model 9303. The warm-up period allows time for the optics to
reach a steady-state temperature. This will result in more repeatable
results. Neglecting the warm-up time may result in an overestimation of
particle size (approximately 10 percent). Neglecting the warm-up time will
not have a significant effect on particle concentration; therefore, a zero
count can still be performed without waiting for the warm-up period.
Main M e n u a n d Basic Op e r at io n
Press Menu on the Default Screen to display the Main Menu.
Main Menu
Main Menu
From the Main Menu you can select other menus:
Set language, time and date, time and day format, and
screen settings.
Set sample mode, cycle mode, sample time, hold time,
cycle count, sample units, and channel 2 size.
View the number of records stored in the Model 9303,
display buffer records, and clear the buffer.
Adjust the pump speed, turn data logging on and off, and
select the sampling location.
Each of these menus is described in the remainder of this chapter.