Operation 3-3
Warm - up T im e
An initial warm-up period of 15 minutes is recommended before sampling
with the Model 9303. The warm-up period allows time for the optics to
reach a steady-state temperature. This will result in more repeatable
results. Neglecting the warm-up time may result in an overestimation of
particle size (approximately 10 percent). Neglecting the warm-up time will
not have a significant effect on particle concentration; therefore, a zero
count can still be performed without waiting for the warm-up period.
Main M e n u a n d Basic Op e r at io n
Press Menu on the Default Screen to display the Main Menu.
Main Menu
MAIN MENU
Settings
Sample Menu
Buffer Menu
Utility Menu
Main Menu
From the Main Menu you can select other menus:
Menu
Description
Settings
Set language, time and date, time and day format, and
screen settings.
Settings
Set sample mode, cycle mode, sample time, hold time,
cycle count, sample units, and channel 2 size.
Buffer
View the number of records stored in the Model 9303,
display buffer records, and clear the buffer.
Utility
Adjust the pump speed, turn data logging on and off, and
select the sampling location.
Each of these menus is described in the remainder of this chapter.

Other manuals for TSI Instruments Aerotrak 9303

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments Aerotrak 9303 and is the answer not in the manual?

TSI Instruments Aerotrak 9303 Specifications

General IconGeneral
Size Range .0.3 to 25 µm
Channel Sizes .0.3 µm and 5.0 µm fixed; 0.5, 1.0, 2.0, or 2.5 µm selectable middle channel.
Counting Efficiency .50% at 0.3 µm; 100% for particles > 0.45 µm (per ISO 21501-4 and JIS)
Concentration Limits .2, 000, 000 particles/ft3 at 5% coincidence loss
Zero Count Level .<1 count / 5 minutes (per ISO 21501-4 and JIS B9921)
Flow Rate .0.1 CFM (2.83 L/min)
Calibration .NIST traceable with TSI® calibration system
Sampling Modes .Manual and automatic
Sampling Time .1 second to 99 minutes 59 seconds
Sampling Frequency .1 to 999 cycles or continuous
Light Source .Long life laser diode
Sample Probe/Tubing .Isokinetic sampling probe
Exhaust .Internal HEPA filter
Vacuum Source .Internal pump
External Surface .High impact injection molded plastic
Communication Mode .USB serial output
Data Storage .1, 500 sample records
Status Indicators .Low battery and service
Display .3.2-in. (8.1 cm) 160 x 160 monochrome LCD
AC Power . (power to AC adapter)110 to 240 VAC 50-60 Hz Universal in-line power supply with country specific plugs
DC Power . (power to instrument)9 VDC @ 1.5 A
Battery .Rechargeable Li-Ion
Battery Life .Up to 4.5 hours of continuous use
Recharge Time .< 3 hours
Weight and Dimensions IconWeight and Dimensions
Dimensions (L x W x H)9.1 x 6.3 x 2.4 in. (23 x 11 x 6.2 cm) (without isokinetic inlet)
Weight .1.3 lbs (0.58 kg) with battery