A-1
A P P E N D I X A
Specifications
All specifications meet or exceed JIS B9921. They are subject to change
without notice.
Size Range ....................
Channel Sizes ...............
0.3 µm and 5.0 µm fixed; 0.5, 1.0, 2.0, or 2.5 µm
selectable middle channel.
Counting Efficiency .....
50% at 0.3 µm; 100% for particles > 0.45 µm
(per ISO 21501-4 and JIS)
2,000,000 particles/ft
3
at 5% coincidence loss
Light Source .................
Zero Count Level ..........
<1 count / 5 minutes (per ISO 21501-4 and
JIS B9921)
Flow Rate ......................
Calibration ....................
NIST traceable with TSI
®
calibration system
Isokinetic sampling probe
Sampling Modes ..........
Sampling Time .............
1 second to 99 minutes 59 seconds
1 to 999 cycles or continuous
Exhaust ........................
Vacuum Source ............
Data Storage .................
Status Indicators ..........
Display ..........................
3.2-in. (8.1 cm) 160 x 160 monochrome LCD
External Surface ..........
High impact injection molded plastic
AC Power ......................
(power to AC adapter)
110 to 240 VAC 50-60 Hz Universal in-line
power supply with country specific plugs
DC Power ......................
(power to instrument)
Battery ..........................
Battery Life ...................
Up to 4.5 hours of continuous use
Recharge Time .............
9.1 x 6.3 x 2.4 in. (23 x 11 x 6.2 cm) (without
isokinetic inlet)
Weight ...........................
1.3 lbs (0.58 kg) with battery