3-8 AeroTrak
®
Portable Airborne Particle Counter
Zone Setup Screen
“Zones” are a convenient way to group sample data for printing and
export, and are required for creating standards-based classification
reports. A Zone contains 1 or more “Locations;” this is modeled after
cleanroom standards that prescribe the classification of a zone (or room)
by taking samples at various locations within the zone.
Use the Zone Setup screen to add, delete or edit Zone configurations.
The Zone configuration screen provides the following information for each
zone that is configured.
The name you assigned to the zone.
The classification standard to use for the samples
taken in the Zone. Options include ISO 14644-1,
EU-GMP, Fed Std 209E F, Fed Std 209E, and
None. Use “None” for taking measurements that
are not associated with standards classification.
The Class selected for the classification of the
Zone. Options vary by the Standard selected.
The occupancy status of the Zone. Options vary
based on the selected standard, but include At
Rest, Operational and As Built.
The direction of air flow through the zone: options
are Unidirectional or Multidirectional.
The area of the zone in ft
2
or m
2
.
Largest Particle Size to
Consider
The largest particle size to consider for
classification measurements. Used by most
standards to calculate minimum required sample
volume.
The number of locations within the zone.
The recipe assigned to the zone.