3.1 Horizontal scale
Horizontal scale, also the horizontal time base, is the time value represented by each scale in the horizontal direction,
which is usually expressed as s/div. With the SCALE knob in the horizontal control area, user can adjust the horizontal
scale in 1 – 2 – 5 steps, i.e. 5ns/div, 10ns/div, 20ns/div, 50ns/div……50s/div. Turn clockwise to decrease the scale and
turn counterclockwise to increase, the scale information on the upper left corner of the screen (as shown below) changes
in real time.
Chapter 3 Horizontal System Settings
When changing the horizontal time base, the waveform will expand or compress according to the trigger point position.
3.2 ROLL mode
When the trigger mode is auto, adjust SCALE knob in the horizontal control area to change the horizontal scale to be
slower than 50ms/div, and the oscilloscope will be in ROLL mode and will continuously plot the waveform voltage-time
trend chart on the screen. The earliest waveform first appears on the right end of the screen, and then gradually moves
to the left, as shown below:
Use the slow sweep mode to observe low frequency signal, and setting “channel coupling” method to “DC” is recommended.
Note: “Horizontal displacement”, “extended window”, “protocol decoding”, “pass/fail”, “parameter measurement”,
“waveform recording” and “independent time base” are not available in ROLL mode.
17

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the UNI-T UPO2000E Series and is the answer not in the manual?

UNI-T UPO2000E Series Specifications

General IconGeneral
BrandUNI-T
ModelUPO2000E Series
CategoryTest Equipment
LanguageEnglish

Related product manuals