Figure 35. Temperature Setup
Figure 36. Temperature Calibration Setup
Press ENTER to accept the entry
and return to temperature units.
Press ENTER to return
to TEMPERATURE
Press ENTER after step
6 to accept the entry and
return to CALIBRATE/
REFERENCE menu.
(FIGURE 35)
FAHRENHEIT
CELSIUS
-
+
CALIBRATE TEMP °U TT.T
REFERENCE TEMP °U TT.T
-
+
1. Fill the bottom (about 2-3 inches [100 mm])
of a one quart [1 litre] container with ice.
2. Center the temperature probe in the
container, then put ice all around it and on
top of it until the container is full.
3. Fill the container with water until it is level with the ice.
4. Verification should be performed in still air and
out ot direct sunlight.
5. Allow 2-5 minutes for the temperature measuring
devices to stabilize.
6. Select 0°C [32°F] for the CALIBRATE TEMP by pressing
the up/down buttons to scroll through the numbers and
NEXT to move from one digit of the CALIBRATE
TEMP to the next.
Enter the REFERENCE TEMP
15°C (60°F). Press up/down arrows to
scroll through the numbers, and
NEXT to move from one digit of the
REFERENCE TEMP to the next.