Zero point unstable
General Instruction Manual
Troubleshooting
IM 01U10B00-00EN-R, 3rd edition, 2018-07-09
81 / 90
10.2 Zero point unstable
Malfunction Possible causes Remedy
Zero point
unstable
Measuring tube not
completely filled with
fluid
Check that the measuring tube in the
sensor is completely filled with fluid.
Correct installation, see [}21].
Bubbles or solids in
the fluid
Check pipe and sensor installation,
see [}27].
Correct installation, see [}21].
No electrical
grounding
Ground transmitter and sensor,
see [}44], and [}63].
Check correct connection of connecting
cable shield on transmitter.
Flow meter installed in
proximity to facilities
with strong electro-
magnetic field
Ground transmitter and sensor,
see [}44], and [}63].
Install flow meter as far away as possible
from these electric devices.
Mechanical strain
from traction or
pressure
Eliminate cause for mechanical tension.
Terminal board or
connection terminals
of transmitter or
sensor soiled or damp
Clean terminal board and connection
terminals.
Clean transmitter and/or sensor.
Dry transmitter and/or sensor.
Seal transmitter and/or sensor tightly.
Influence of external
vibration
Install mechanical dampers.
Increase parameter mass flow damping
(see applicable software instruction
manual).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA Rotamass series and is the answer not in the manual?

YOKOGAWA Rotamass series Specifications

General IconGeneral
BrandYOKOGAWA
ModelRotamass series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals