322 AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B
Status messages
List layout
The status message list contains the following information:
No. The status message number appears in the menu line display
Text Full text of the status message is shown in the detailed display
S x = Status message sets the overall status
A x = Status message sets the "Error" individual status
W x = Status message sets the "Maintenance request" individual
status
F x = Status message sets the "Maintenance mode" individual
status
Reaction/
Remark
Explanations and corrective measures in case of status mes-
sages
Status messages
No. Text S A
W
F Reaction/Remark
Run time error
1–
21
Runtime error 1 to
Runtime error 21
Notify Service if these messages occur repeatedly.
System controller
101 System controller shut down at For information; shows date and time
102 System controller starts up at For information; shows date and time as well as warm or
cold start
103 Install Module: For information
104 Remove Module: For information
105 Reactivate Module: For information
106 A user installed module: For information
107 A user removed module: For information
108 A user replaced module: For information
109 A password is active! To delete,
press the <MEAS> key on the
measurement value display.
For information on Password Protection see "Password
protection" (see page 149) section; not logged
110 System booting. Not logged
111 This system is currently under
remote control
Not logged
112 Display/control unit synchro-
nizing with analyzer. Please
wait.
Not logged
113 The system time was changed
from -> to:
Not logged
114 The system is saving the
changed parameters. Please
wait.