EasyManuals Logo
Home>ABB>Measuring Instruments>AO2040

ABB AO2040 User Manual

ABB AO2040
362 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #351 background imageLoading...
Page #351 background image
AO2000 CONTINUOUS GAS ANALYZERS | OI/AO2000-EN REV. B 351
Magnos206: Operating specifications
Linearity deviation ≤ 0.5 % of span
Repeatability ≤ 50 ppm O
2
(time base for gas exchange ≥ 5 minutes)
Zero drift ≤ 3 % of span of the smallest measurement range (per order) per week,
minimum 300 ppm O
2
per week following prolonged transport and storage
time the drift can be higher during the first weeks of operation.
Sensitivity drift ≤ 0.1 vol.% O
2
per week or ≤ 1 % of measured value per week (not cumula-
tive), the smaller value applies; ≤ 0.25 % of the measured value per year
Output fluctuation (2
σ)
≤ 25 ppm O
2
at electronic T90 time (static/dynamic) = 3/0 s
Detection limit (4
σ)
≤ 50 ppm O
2
at electronic T90 time (static/dynamic)= 3/0 s
Flow influence ≤ 0.1 vol.% O
2
in permissible range
Associated gas effect Influence effect of the associated gases is described in the IEC 61207-3:2002
"Gas analyzers – Expression of performance – Part 3: Paramagnetic oxygen
analyzers"
Temperature influence At zero point: ≤ 0.02 vol.% O
2
per 10 °C;
On sensitivity: ≤ 0.1 % of the measured value per 10 °C.
Thermostat temperature = 64 °C; ambient temperature in permissible range
Air pressure effect At zero point: no influence effect;
On sensitivity without pressure correction: ≤ 1 % of the measured value per
1 % air pressure change;
On sensitivity with pressure correction by means of internal pressure sensor:
≤ 0.1 % of the measured value per 1 % of air pressure change; for highly sup-
pressed measurement ranges ≤ 0.01 % of the measured value per 1 % of air
pressure change or ≤ 0.002 vol.% of O
2
per 1 % pressure change, whichever is
greater.
Pressure sensor working range: p
abs
= 600 to 1250 hPa
Power supply influence 24 V DC ± 5 %: ≤ 0.4 % of span
Position influence Zero-point shift ≤ 0.05 vol.% O
2
per 1° deviation from horizontal orientation.
Position has no effect on the hard-mounted unit.
T
90
time T
90
≤ 3.5 to 10 s at sample gas flow = 10 l/h and electronic T90 time (static/
dynamic) = 3/0 s, gas switching from N
2
to air (applies to a gas analyzer with
one analyzer module)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB AO2040 and is the answer not in the manual?

ABB AO2040 Specifications

General IconGeneral
BrandABB
ModelAO2040
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals