Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers 6 Configuration
34 IM/AZT6CR–EN Rev. J
6.4 Alarm Relays
Fig. 6.3 High/Low Process Alarms
Fig. 6.4 High/Low Latch Alarms
Fig. 6.5 High/Low Annunciate Alarms
Hysteresis
Concentration
Alarm On
Alarm switches
off automatically
Alarm Off
Trip Po int
High Alarm Action
Low Alarm Action
Hysteresis
Concentration
Concentration
Alarm On
Alarm switches
off automatically
Concentration
Hysteresis
Alarm Off
Trip Point
Alarm On Alarm Latched
Alarm
Off
Alarm
acknowledged
by operator
Earliest point that
the alarm can be
acknowledged
Earliest point that
the alarm can be
acknowledged
Hysteresis
Alarm On Alarm Latched
Concentration
Alarm
acknowledged
by operator
High Latch Alarm Action
Low Latch Alarm Action
Concentration
Alarm
acknowledged
by operator
Concentration
Alarm switches
off automatically
Alarm Off
Trip Point
Alarm
acknowledged
by operator
Hysteresis
Alarm On
Alarm switches
off automatically
High Annunciate Alarm Action
Low Annunciate Alarm Action

Table of Contents

Other manuals for ABB Aztec 600

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ABB Aztec 600 and is the answer not in the manual?

ABB Aztec 600 Specifications

General IconGeneral
BrandABB
ModelAztec 600
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals