Chapter 10 147
Command Reference
:CALC1:FORM
10. Command Reference
:CALC1:FORM
Syntax :CALCulate1:FORMat {CP|CS}
:CALCulate1:FORMat?
Description Specifies the primary parameter to be measured.
When the secondary parameter has been set to G or RP and the primary parameter is set to
CS, the secondary parameter is automatically changed to D. Also, when the secondary
parameter has been set to RS and the primary parameter is set to CP, the secondary
parameter is automatically changed to D.
Parameters
Query response {CP|CS}<newline><^END>
Related commands :CALC2:FORM on page 150
Equivalent key
sequence
[Meas Prmtr]
:CALC1:MATH:EXPR:CAT?
Syntax :CALCulate1:MATH:EXPRession:CATalog?
Description Reads out available parameters used when executing the command to specify the
expression of the primary parameter in the deviation measurement mode (the
(:CALC1:MATH:EXPR:NAME command). The query response is always DEV,PCNT.
(Query only)
Query response DEV,PCNT<newline><^END> (fixed)
Related commands :CALC1:MATH:EXPR:NAME on page 148
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
CP (initial value) Specifies the capacitance value measured using the parallel
equivalent circuit model for the primary parameter.
CS Specifies the capacitance value measured using the series
equivalent circuit model for the primary parameter.