Chapter 10 165
Command Reference
:CORR:OPEN
10. Command Reference
:CORR:OPEN
Syntax [:SENSe]:CORRection:OPEN[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:OPEN[:STATe]?
Description Turns ON/OFF the OPEN compensation.
With the OPEN compensation set to ON, if you change the cable length (set with the
:CAL:CABL command) or frequency shift (set with the :SYST:FSH command), the
OPEN compensation is automatically changed to OFF.
Parameters
Query response {1|0}<newline><^END>
Related commands :CAL:CABL on page 133
:SYST:FSH on page 192
:CORR:COLL on page 157
Equivalent key
sequence
[Open]( [Blue] - [4] ) - On/Off
:CORR:SHOR
Syntax [:SENSe]:CORRection:SHORt[:STATe] {ON|OFF|1|0}
[:SENSe]:CORRection:SHORt[:STATe]?
Description Turns ON/OFF the SHORT compensation.
With the SHORT compensation set to ON, if you change the cable length (set with the
:CAL:CABL command) or frequency shift (set with the :SYST:FSH command), the
SHORT compensation is automatically changed to OFF.
Parameters
Query response {1|0}<newline><^END>
Related commands :CAL:CABL on page 133
:SYST:FSH on page 192
:CORR:COLL on page 157
Equivalent key
sequence
[Short]( [Blue] - [5] ) - On/Off
Description
ON or 1 Turns ON the OPEN compensation.
OFF or 0 (initial value) Turns OFF the OPEN compensation.
Description
ON or 1 Turns ON the SHORT compensation.
OFF or 0 (initial value) Turns OFF the SHORT compensation.