164 Chapter 10
Command Reference
:CORR:OFFS:DATA
:CORR:OFFS:DATA
Syntax [:SENSe]:CORRection:OFFSet:DATA <numeric 1>,<numeric 2>
[:SENSe]:CORRection:OFFSet:DATA?
Description Sets the compensation values for the primary parameter and secondary parameter used in
the offset compensation function.
These compensation data are measured for the measurement frequency used in executing
the command (set with the :SOUR:FREQ command).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set. Note that the unit of the parameter changes depending on the
limit range designation method.
MAX or MIN can be used to specify the parameters.
Query response {numeric 1},{numeric 2}<newline><^END>
Related commands :CORR:OFFS on page 163
Equivalent key
sequence
[Offset]( [Blue] - [3] ) - OfsEnt
<numeric 1> <numeric 2>
Description The offset compensation value for
the primary parameter.
The offset compensation value for
the secondary parameter.
Range -999.999 to 999.999 -99.9999E9 to 99.9999E9
Initial value 0 0
Unit F (farad) Depends on the setup of the
secondary parameter.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 4288A and is the answer not in the manual?

Agilent Technologies 4288A Specifications

General IconGeneral
BrandAgilent Technologies
Model4288A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals