Chapter 10 195
Command Reference
:TRIG
10. Command Reference
:TRIG
Syntax :TRIGger[:SEQuence1][:IMMediate]
Description Immediately generates a trigger and executes a measurement regardless of the setup of the
trigger mode.
If the trigger system is not in the trigger wait state (trigger event detection state), executing
this command causes an error and the command is ignored.
For details on the trigger system, refer to “Trigger system” on page 52. (No query)
Equivalent key
sequence
No equivalent key is available on the front panel.
:TRIG:DEL
Syntax :TRIGger[:SEQuence1]:DELay <numeric>[MS|M|S]
:TRIGger[:SEQuence1]:DELay?
Description Specifies the waiting time between when a trigger is inputted and when the measurement
starts (trigger delay time).
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set. Note that a fraction below the resolution is rounded off.
MAX or MIN can be used to specify the parameters.
Query response {numeric}<newline><^END>
Equivalent key
sequence
[Delay]( [Blue] - [Trig Mode] )
<numeric>
Description Trigger delay time
Range 0 to 1
Initial value 0
Unit s (second)
Resolution 1E-3