EasyManuals Logo
Home>Agilent Technologies>Test Equipment>8960 Series

Agilent Technologies 8960 Series User Manual

Agilent Technologies 8960 Series
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #116 background imageLoading...
Page #116 background image
116
S:\Hp8960\Generic Documents\Manual Operation Getting Started Guide\Pi_manual operation getting started guide\chapters\gprs_man_measurements.fm
GPRS Mobile Measurements
Measuring Output RF Spectrum
1. Connect the mobile.
2. Turn the mobile on and wait for Attached in the Active Cell: field.
Note: For mobiles that dont perform a GPRS attach automatically, set the mobile
to data mode.
3. Press the Start Data Connection (
F3) key and watch for the Active Cell:
field changing to Transferring.
4. Press the Measurement selection key.
5. Select the Output RF Spectrum measurement.
6. Press the ORFS Setup (
F1) key.
7. Press the Measurement Setup (
F1) key.
8. Set the measurement parameters as needed for your measurement, including:
Measurement Timeout = 20 s.
9. Press the Modulation Frequencies (
F2) key and set up your required offset
frequencies for ORFS due to modulation.
10.Press the Switching Frequencies (
F3) key and set up your required offset
frequencies for ORFS due to switching.
11.Press the Close Menu (F6) key.
12.Press the Return (
F6) key.
13.Press the Change View (
F2) key.
14.Press the Modulation Numeric 1 of 3 (
F1) key.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals