cdma2000 Measurement Procedures
Measuring Access Probe Power
1. Press the CALL SETUP key.
2. Press the End Call key (
F3).
3. Select the Access Probe Power measurement (see “B. Select measurements.” on
page 48).
4. Press the Register Mobile key (
F6). A typical display is shown below.
NOTE The remainder of this procedure lists the steps required to make access
probe power measurements using call limit mode. Call limit mode
provides testing of complete access probe sequences during a call
origination.
5. Set up the measurement parameters as needed for your measurement situation
(see “B. Set up the measurement.” on page 50).
6. Press the
More key on the lower left side of the display (Call Control) until 2 of 2 is
displayed. Press the Cell Info key (
F2) key, then the Registration
Parameters key.
7. Turn the Timer Based Registration State off.
8. Press the
MEASUREMENT key.
9. Press the
More key on the lower right side of the display (Call Parms) until 2 of 3 is
displayed. Press the Call Limit Mode key (
F9) key and turn call limiting on.