cdma2000 Measurement Procedures
Measuring Handoff Waveform Quality
Each waveform quality measurement returns a number of measurements which
includes rho.
These measurements can be made only when one of the following radio
configurations (see “How Do I Set Up a Call?” on page 52) is selected:
• (Fwd3, Rvs3)
• (Fwd4, Rvs3)
1. Select the Handoff Waveform Quality measurement (see “B. Select
measurements.” on page 48).
2. Set up the measurement parameters (Measurement Timeout is the only
parameter provided with this measurement) as needed for your measurement
situation (see “B. Set up the measurement.” on page 50.)
3. Press the
START SINGLE key. A typical display is shown below.
In addition to rho, the waveform quality measurement also provides the following
measurements, calculated from the same data:
• Time Error
• Carrier Feedthrough
• Phase Error
• Magnitude Error