cdma2000 Measurement Procedures
Measuring Waveform Quality
Each waveform quality measurement returns a number of measurements which
includes rho.
1. Select the Waveform Quality/Code Domain measurement (see B. Select
measurements. on page 48). A typical display is shown below.
2. Set up the measurement parameters as needed for your measurement situation
(see B. Set up the measurement. on page 50.
Numeric rho is the screen currently displayed. In addition to rho, it also provides the
following measurements, calculated from the same data:
Time Error
Carrier Feedthrough
Phase Error
Magnitude Error
EVM
There are graphical measurement results available as a result of performing this
measurement. See Measuring Code Domain Power and Code Domain Power + Noise
on page 79.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals