BPAP Series User Manual V1.0
43 / 58
(2) Press the home button and it will become silent for 100 to 120 seconds. If the
alarm state has not been eliminated, and then the alarm sounds again immediately after the
end of the silence.
(3) Turn the button leftwards or rightwards to reduce the residual alarming information.
Low minute ventilation alarm test
(1) Connect the device to the simulated lung.
(2) Observe the value of minute ventilation displayed on the screen.
(3) Make the alarm value of the minute ventilation larger than the displayed value, and then
confirm whether the alarm of low minute ventilation occurs.
(4) Press the home button and it will become silent for 100 to 120 seconds. If the
alarm state has not been eliminated, and then the alarm sounds again immediately after the
end of the silence.
(5) Turn the button leftwards or rightwards to reduce the residual alarming information.
(6) Set the alarm setting of the low minute ventilation as "Off".
Power failure alarm test
(1) Confirm whether an audible alarm will sound in 6s when the device is accidentally
disconnected from power during it is delivering air.
(2) Reconnect the power supply, and then confirm if the device restarts delivering air.
WARNING!
Adjust the device to the appropriate patient settings after the test and before use.
18. Cleaning and Disinfection
WARNINGS!
Regular cleaning of the device and its accessories is very important for the prevention of
respiratory infections.
• To avoid electric shock, always unplug the device before cleaning.
Use mild soap that is nontoxic to humans.
Follow the manufacturer's instructions on cleaning the mask and tubing and on determining
the frequency of cleaning.
• Before cleaning, check whether the device has been disconnected from the power supply,
whether the power cord has been unplugged, and whether the water chamber of the device
has cooled down. Make sure the plate has cooled down to room temperature, so you do not
get burned.
• Do not open or modify the device. There are no user serviceable parts inside. Repairs and
servicing should only be performed by an authorized service agent.
CAUTIONS!
Overheating of the materials could lead to early fatigue of these materials.