G2S BPAP Series User Manual V2.1
6 / 43
Sound Pressure Level
< 30 dB, when the device is working at the pressure of 10 hPa.
Sound Power Level
< 38 dB, when the device is working at the pressure of 10 hPa.
Maximum Flow
For models with a maximum pressure of 20 hPa, the flow table is as follows.
Test Pressure (hPa)
4
8
12
16
Measured Pressure at the Patient
Connection Port (hPa)
3
7
11
15
Average Flow at the Patient
Connection Port (L/min)
90
150
150
150
When the working pressure is set to the values listed in the table, the average flow rate at
the patient end should be greater than 80% of the corresponding flow value in the table.
For models with a maximum pressure of 25 hPa, the flow table is as follows.
Test Pressure (hPa)
4
10
15
20
Measured Pressure at the Patient
Connection Port (hPa)
3
9
14
19
Average Flow at the Patient
Connection Port (L/min)
90
150
150
145
When the working pressure is set to the values listed in the table, the average flow rate at
the patient end should be greater than 80% of the corresponding flow value in the table.
For models with a maximum pressure of 30 hPa, the flow table is as follows.
Test Pressure (hPa)
4
11
17
24
Measured Pressure at the Patient
Connection Port (hPa)
3
10
16
23
Average Flow at the Patient
Connection Port (L/min)
90
150
150
120
When the working pressure is set to the values listed in the table, the average flow rate at
the patient end should be greater than 80% of the corresponding flow value in the table.
SpO
2
Range: 0 100%
The margin of error for SpO
2
between 70% and 100% is ± 3%. No strict accuracy
requirements for SpO
2
below 70%.
Pulse Rate
Range: 40 240 BPM
Margin of Error: ± 1%
Wavelengths
Red: 663 nanometers
Infrared: 890 nanometers
Maximal Optical Output Power
Less than 1.5 mW maximum average.

Table of Contents

Other manuals for BMC G2S BPAP Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BMC G2S BPAP Series and is the answer not in the manual?

BMC G2S BPAP Series Specifications

General IconGeneral
BrandBMC
ModelG2S BPAP Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals