CR:800C Series Sound Level Meter User Manual
Page 31
In this example, the measurement duration will be set to 15 Minutes, and the instrument will
run continuously until the Stop key is pressed. The change the measurement duration, use the
Up and Down keys to select the require duration and press OK to select.
In addition to the preset measurement durations, the user can define the measurement duration
using the Set User Time Option.
Use the Up and Down keys to change the highlighted parameter and the Left and Right keys to
move through the different fields. Press OK to accept the User Time.
4.5 Measurement Auto Repeat
The auto-repeat function is used to set the instrument to make a series of contiguous
measurements. For example, if the measurement duration was set to 15 minutes and
the Auto Repeat disabled, after one 15 minute measurement the instrument will stop
and store the measurement information the memory.
Please note that this function only operates in the Broadband Measurement. Auto
Repeat is not available during 1:1 or 1:3 Octave Band Mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cirrus Research CR:811C and is the answer not in the manual?

Cirrus Research CR:811C Specifications

General IconGeneral
BrandCirrus Research
ModelCR:811C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals