14 Dräger X-dock 5300/6300/6600
Basics
4.9 Manage tests
Select > Define test.
The test overview will be displayed.
4.9.1 Selecting the test mode
The following test modes are available. The test modes define
the behaviour of the system when the user is logged off. The
test mode Test scheduler can be configured.
To select the test mode:
1. Select .
2. Select desired test mode.
3. If necessary, select of the test mode to be edited (only
possible with the Test scheduler test mode).
4. If necessary, edit tests (see chapter 4.9.2 on page 14).
5. Save settings with OK.
4.9.2 Creating or editing a test
To create or edit a test:
1. Select the symbol of the test to be edited.
The editing window is displayed.
2. Enter desired test name.
3. Select the desired gas test.
The following gas tests are available:
Tests with measured gas
Select
test mode
►◄
Configure
test mode
Select test 1 ►◄Edit test 1
Select test 2 ►◄Edit test 2
Select test 3 ►◄Edit test 3
Mode
Description
Favorites
mode
The Favorites mode displays the predefined tests.
These can be selected via the favourites bar.
Single mode In Single mode, a predefined test is started by
closing the module flap. Several tests can be
started and executed in parallel.
Test
scheduler
The Test scheduler enables the user to
configure the time and weekday when a
predefined test should be performed. To
configure the Test scheduler mode, select .
Manual tests can only be started when the user
is logged in.
NOTICE
A maximum of 3 favourites may be created.
The first 3 characters of the test name are displayed in
the list of favourites.
Enter test name ►
Select gas test ►
Select permission level ►
Options ►
Further options ►
-
No gas test, only the activated
options are run.
Quick bump test
Test threshold override test
(if test gas = measured gas,
concentration A1; see chapter 7. on
page 15).
Extended bump test
Test for reaching cylinder
concentration within a tolerance
range (see chapter 7. on page 15).
Calibration
Zero point and sensitivity
adjustment
i
i