EasyManuals Logo
Home>Dräger>Test Equipment>drager X-dock 5300

Dräger drager X-dock 5300 Technical Manual

Dräger drager X-dock 5300
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
Basics
Dräger X-dock 5300/6300/6600 15
Tests with substitute gas
1
4. Assign the desired authorisation level.
5. Select Options if necessary.
The following options are available:
6. If required, select Further options:
7. Save settings with OK.
4.10 Set up template configuration
The maintenance station can configure gas monitors
automatically using a configuration template. The
configuration templates can be created using the Dräger CC-
Vision PC software. The configuration templates can be
transferred to the maintenance station using a USB memory
stick or the X-dock Manager.
1. Create a configuration template using the CC-Vision PC
software.
2. Transfer the template file to the station using a USB
memory stick or the X-dock Manager (see chapter 6.5.7 on
page 21 and online help for X-dock Manager).
3. When defining a test, select the desired option:
a. Select Check config. whether to carry out just one
check of the configuration.
-
No gas test, only the activated options are
run.
Schneller
Begasungst
est
For tests with substitute gas, a switch is
made to the test gas. So, for example, the
influence of temperature is reduced and a
greater accuracy is achieved.
Since the alarm thresholds no longer apply
here, it is always tested at a calculated
threshold. The threshold is calculated at 1/
3*test gas concentration.
Erweiterter
Begasungst
est
For tests with substitute gas, a switch is
made to the test gas. So, for example, the
influence of temperature is reduced and a
greater accuracy is achieved.
For the extended test, a check is carried out
on the test gas concentration with the set
tolerance.
Calibration
For calibration with substitute gas, a switch
is made to the test gas. So, for example, the
influence of temperature is reduced and a
greater accuracy is achieved.
Calibration uses the test gas concentration
of the test gas. Following successful
calibration, a switch is made to the
measured gas.
1 (The Pac 7000 OV PAC 7000 OV-A does not support switching to
the test gas. For this reason, the X-dock performs the conversion
here. There is no compensation for the effects of temperature, so
testing/calibration is only recommended in a range of from 0-
30 °C.)
Response time test
(only for calibration)
The response times of the sensors
are tested.
Alarm test
All the alarm functions of the gas
monitor are checked (noise, light,
vibration if configured).
Zero check
The zero point is checked.
Check config.
The configuration of the gas monitor
is compared with the template
activated at the station. If the
configuration does not match, an
error message is generated.
Set config.
If this option is activated, the gas
monitors are configured according
to the activated template at the
station if a deviation has previously
been determined.
NOTICE
When the alarm functions are being tested, the
ambient noise must not be too loud as otherwise the
station will not be able to check the horn function.
i
i
Time sync.
The time on the gas monitor is
synchronised with the station.
If this option is selected, Reset DL
and Reset TWA are automatically
also activated to ensure that the DL
(data logger) is unique.
Download DL
Download data memory and events
memory data from the gas monitor
to the station.
Reset DL
The data in the data and events
memories are cleared.
Reset TWA
The TWA
1
time is reset.
1 Average shift values (time-weighted average) are generally
limited to eight hours exposure per day per workplace for
5 days a week during a work lifetime.
Turn off
The gas monitor is switched off
automatically after the test.
AutoCal
Perform automatic calibration (e. g.
calibrate after a faulty bump test).
Purge
The system is purged with fresh air
after every test.
Generate
2
2 In the case of networked systems, Dräger recommends that
you do not activate the “Generate” option but create the
certificate using the X-dock Manager. This reduces the test
duration.
Create certificate (stored in the
station as a PDF file).
Print
Print out the certificate on a
connected USB printer.
NOTICE
Before changing the configuration, Dräger
recommends:
1. Transferring the new configuration template to a
gas monitor using an X-dock station.
2. Checking the new configuration on the gas monitor
using the Dräger CC-Vision PC software,
3. approving the configuration template and
distributing it to the corresponding X-dock station.
i
i

Other manuals for Dräger drager X-dock 5300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger drager X-dock 5300 and is the answer not in the manual?

Dräger drager X-dock 5300 Specifications

General IconGeneral
BrandDräger
Modeldrager X-dock 5300
CategoryTest Equipment
LanguageEnglish

Related product manuals