Troubleshooting
14 Tips for operators and service technicians
Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our
service.
WARNING
Infection due to contaminated unit
Clean and disinfect the suction before working on the unit.
Wear protective equipment when working (e. g. impermeable gloves, protective goggles and
mouth and nose protection).
Prior to working on the unit or in case of danger, disconnect it from the mains.
Error Possible cause Remedy
Device does not start
No mains voltage
Check the mains voltage. *
Check the fuses and replace if
necessary. *
Undervoltage
Measure the supply voltage;
call an electrician if necessary.
*
No start signal
Check the control voltage at
the signal input. *
Capacitor defective
Measure capacitance and
replace if necessary. *
Turbine is blocked by solid parti-
cles or sticky soiling
Disassemble the unit and
clean the turbine. *
The unit generates unusual
noises
Solid particles in the turbine
chamber
Disassemble the unit and
clean the turbine and housing.
*
Water leaking from the
exhaust air connection
Membrane valve blocked
Check the membrane valve at
the waste water connection
and if necessary clean or
replace. *
Foam in turbine due to use of
incorrect cleaning and disinfec-
tant agents
Use non-foaming cleaning
and disinfectant agents.
Build-up of condensate in the
exhaust air line
Check the pipe system; avoid
over-cooling. *
Waste water line/siphon trap
blocked
Clean the waste water line/
siphon trap. *
Troubleshooting
7119100007L02 1909V001 35
EN

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Durr Dental V 300 S and is the answer not in the manual?

Durr Dental V 300 S Specifications

General IconGeneral
BrandDurr Dental
ModelV 300 S
CategoryMedical Equipment
LanguageEnglish

Related product manuals