FEASA LED ANALYSER
FUNCTIONAL VERSION
User Calibration Mode
setCALIBRATION## - Select the Calibration Set to use
Transmit Receive
setcalibration##
putcalibration##
OK
OK
Where:
## represents the set value (## can be 00 to 50).
Description
This command will select which calibration set you are accessing. The Led analyser is capable
of storing fifty one (51) separate user Calibration sets on the board.
The command SETCALIBRATION## stores the Calibration sets both on RAM and on EEPROM
(stays after power cycle but limits the number of set command operations to ~ 100K).
The command PUTCALIBRATION## only stores them on RAM and the analyser will default
back to what values were stored in the EEPROM after a power cycle.
The default calibration set is 00
Example:
If you want to select set No.3
setcalibration03
OK
Example:
If you want to select set No.3
putcalibration03
OK
The command GETCALIBRATIONSET will return the number of which stored Calibration set of
values are currently selected.
Example:
calibrationset03
Back to Index