12 Specifications and Operation DATAPAQ TP3
Ten or 20 thermocouple channels (depending on model size; see below)
for maximum data collection on each run.
Can be specified for use with various thermocouple types (B, J, K, N, R,
S, T; see p. 14) and with analog inputs (current and/or voltage; see
p. 29) – or with a mixture of thermocouple types and/or analog inputs.
Powered by standard alkaline AA batteries or by user-replaceable
rechargeable NiMH batteries; for high-temperature applications
involving logger operating temperatures (i.e. inside the thermal barrier) of up
to 110°C/23F, can also be powered by non-rechargeable lithium
batteries. Any logger can use any battery-type interchangeably if
appropriate battery housings are used (available separately). See p. 17.
Short sample intervals, to collect maximum data in minimum time
(p. 13).
High accuracy for compliance to tight specifications (see p. 14):
for type K thermocouples, ±0.3°C/0.5°F (above −100°C/148°F);
for type N thermocouples, ±0.4°C/0.7°F (above 0°C/32°F).
Data gathered by the logger but not yet downloaded (hot data) is
protected by non-volatile memory or by software warning if reset is
attempted before download.
USB and Bluetooth communication (p. 29).
Hardwired telemetry (p. 55) or (if specified for use with optional
TM21 system) radio telemetry for monitoring in real time – both with full
analysis functions and alarms to warn the user if the process is out of
specification.
Four LEDs to show the exact status of the loggers activity and its batteries
(p. 15).
Start and stop buttons for easy user control (p. 17, p. 53).
Rugged case and electronics allow operation in harsh environments of
dust, pressure and vacuum.
Reset possible by start button alone, for speed and simplicity (p. 47).
Specifications
Narrow – 10-channel
TP3000A
Narrow – 20-channel
TP3300A
Wide – 20-channel
TP3400A
Height
Width
Length
20.5 mm/0.8 in.
97 mm/3.8 in.
198 mm/7.8 in.
20.5 mm/0.8 in.
97 mm/3.8 in.
198 mm/7.8 in.
20.5 mm/0.8 in.
124 mm/4.9 in.
177 mm/7.0 in.