Appendix B
2008K Hemodialysis Machine Operator’s Manual Rev P 209
Auto Heat Disinfect
The Auto Heat Disinfect option is a special feature* that allows the user to program the
2008K hemodialysis machine to automatically run a heat disinfection program according to a
schedule. In the “Options: Default Settings” screen the user may select a start time (Start
Prg) and any day(s) of the week toggle-button(s).
This program affects only the start time of the Heat Disinfect, all other settings function as
usual. With the machine on and set up for rinse, the Heat Disinfect program will run at the
selected time(s).
Note: If another rinse is running when the Auto Heat Disinfect is set to run, the Auto Heat
Disinfect will start after the first rinse completes.
Setting Auto Heat Disinfect Times
1. Go to Service Mode and select the “Options: Default Settings” screen (see below).
Figure 72 – Service Mode: Options: Default Settings Screen
* The Auto Heat Disinfection feature is only available with kit P/N 190679. Contact the
Fresenius Medical Care Spare Parts department or your sales representative for more information
and pricing.
2) Select Auto
Disinfect to
display the “Auto
Disinfect” screen.

Table of Contents

Other manuals for Fresenius Medical Care 2008K

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008K and is the answer not in the manual?

Fresenius Medical Care 2008K Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008K
CategoryMedical Equipment
LanguageEnglish

Related product manuals