Appendix B
216 2008K Hemodialysis Machine Operator’s Manual Rev P
Equipment Storage and Maintenance
Follow the storage and maintenance procedures for dialysis equipment used for Intermittent
Hemodialysis (IHD) in the ICU setting.
Warning! Possible explosion hazard if used in the presence of flammable anesthetics.
Storage Location
If the dialysis unit is incorporated in the hospital, the equipment should be stored so that it
will not be damaged. The maintenance of the equipment is normally part of the duty of the
dialysis service technician. Depending on the disinfectant and the storage time, frequent
flushing of the equipment is necessary.
If it is more convenient to store the equipment close to the ICU, the dedicated storage space
should have an access to tap water, power and a drain. The room should be well vented, if
disinfectant is used.
Storage Preparation
Before storing the 2008K hemodialysis machine, the hydraulics should be disinfected. It is
also necessary to wipe the external parts of the machine with a surface cleaner. Frequency of
disinfection and length of the dwell time depend on the disinfectant and shall be determined
by acceptable culture result. The following table lists commonly used procedures to store
equipment and then return it from storage. Validate your own procedure according to the
facility’s policy. After prolonged storage, use a bleach disinfectant prior to patient use.
Table 34 – Disinfectant Information
Disinfectant Program Dwell time Retest and Repeat
Formaldehyde
Fill Hydraulic
(Service Mode)
Unlimited 3 – 4 weeks
Bleach
Chemical Disinfection
program
Only for time of
disinfection program
24 hours
Renalin
Chemical Disinfection
program
Only for time of
disinfection program
24 hours
Heat
Heat Disinfection
program
Recirculate and shut
machine off
24 hours

Table of Contents

Other manuals for Fresenius Medical Care 2008K

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008K and is the answer not in the manual?

Fresenius Medical Care 2008K Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008K
CategoryMedical Equipment
LanguageEnglish

Related product manuals