Chapter 2—Daily Preparation for Treatment
2008K Hemodialysis Machine Operator’s Manual Rev P 53
To run the Pressure and Alarms tests:
Figure 23 – Starting Automatic Tests
Test Sequence
The automated test sequence consists of two distinct parts—Alarm tests and Pressure Holding
tests. The Pressure Holding Test, the Alarm test, or both tests can be started by touching the
1) From any
screen, touch
Test & Options
3) Press
CONFIRM
2) From the
“Test & Options”
screen, touch
Both Tests