2-16 Fresenius Medical Care 4008 4/09.03 (TM)
4 Ultrafiltration system and membrane pumps
TSC 4.1 UF pump, 1 stroke = 1 ml, 60 strokes = 60 ml ±0.5 ml........... Measured value:________ ❏
MA 4.2 Average balancing chamber volume 30 ± 1 ml....................... Measured value:________ ❏
MA 4.3 Concentrate pump calibration volume removal /
number of strokes ................................................................... Measured value:________ ❏
MA 4.4 Bicarbonate pump calibration volume removal /
number of strokes ................................................................... Measured value:________ ❏
5 Dialysis mode
MA 5.1 Expected temperature 37 °C ± 0.5 °C .................................... Measured value:________ ❏
MA 5.2 Temperature display 37 °C ± 0.5 °C ....................................... Measured value:________ ❏
MA 5.3 Dialysate flow check
800 ml/min (desired value: 765 to 837 ml/min) ................... Measured value:________ ❏
500 ml/min (desired value: 471 to 528 ml/min) ................... Measured value:________ ❏
300 ml/min (desired value: 279 to 321 ml/min) ................... Measured value:________ ❏
MA 5.4 Dialysate pressure
– Check zero point with flow off ........................................................................................ ❏
– Slope checked ................................................................................................................ ❏
TSC 5.5 Conductivity display checked with reference meter........................................................... ❏
If the bibag
®
option is used, connect a bibag
®
!
– CD system ...........................................................................Measured value:________ ❏
– CD ref...................................................................................Measured value:________ ❏
6 Extracorporeal components
MA 6.1 Arterial pressure displayed checked with reference meter ................................................ ❏
MA 6.2 Venous pressure displayed checked with reference meter ............................................... ❏
TSC 6.3 Blood pumps: blood pump rate checked (calibration program: BP-Rate CHECK)............ ❏
TSC 6.4 SN switching pressure checked according to table in TM ................................................. ❏
TSC 6.5 Blood pump stop alarm checked ....................................................................................... ❏
TSC 6.6 Venous line clamp closes after blood alarm ...................................................................... ❏
TSC 6.7 Pressure of about 2 bar in the venous bubble catcher ...................................................... ❏
Pressure must not drop by more than 0.1 bar within 3 minutes. ....................................... ❏
7 Options
7.1 bibag
®
MA 7.1.1 bibag
®
connector, O-rings replaced ................................................................................... ❏
MA 7.1.2 Switching pressure of PSW134 checked,
100 mbar, + 10 mbar............................................................... Measured value:________ ❏
7.2 DIASAFE
MA 7.2.1 DIASAFE filter life checked................................................................................................ ❏
MA 7.2.2 Hydrophobic filter 111 replaced ......................................................................................... ❏
MA 7.2.3 O-rings in the dialysate couplings of the DIASAFE replaced ............................................ ❏
7.3 DIASAFE
®
plus
MA 7.3.1 DIASAFE
®
plus
filter life checked ......................................................................................... ❏
MA 7.3.2 Hydrophobic filter 111 replaced ......................................................................................... ❏
7.4 4008 HDF
TSC 7.4.1 2
nd
UF pump 1 stroke = 1 ml, 60 strokes = 60 ml ± 0.5 ml ..... Measured value:________ ❏
7.5 ON-LINE-HDF (and DIASAFE)
MA 7.5.1 Filter life of the DIASAFE and ON-LINE filter checked...................................................... ❏
MA 7.5.2 Hydrophobic filter 111 replaced ......................................................................................... ❏
MA 7.5.3 O-rings in the dialysate couplings of the DIASAFE replaced ............................................ ❏
MA 7.5.4 HDF pump rotor checked (smooth running, wear) ............................................................ ❏
MA 7.5.5 Fastening strap for Luer-lock checked .............................................................................. ❏
TSC 7.5.6 Substituate pump ................................................................................................................ ❏
❏ (part no. 672 521 1) with DC motor:
volume delivered by the pump checked ............................... desired/actual:____/____
or
❏ (part no. 674 982 1) with stepper motor:
pump rate checked (calibration program: HDF-P.-Rate CHECK)