Fresenius Medical Care 4008 4/08.03 (TM) 1-7
1.1.3 Description of the T1 test incl. error messages
Notes on the descriptions of errors
The values specified are internal mathematical values, which are used in the program.
Resolutions and tolerances are as follows:
1. Arterial pressure: 3 mmHg/digit and ± 1 digit of measured value.
2. Venous pressure: 3 mmHg/digit and ± 1 digit of measured value.
3. Dialysate pressure, coarse: 6.0 mmHg/digit and ± 1 digit of measured value.
fine: 0.5 mmHg/digit and ± 1 digit of measured value.
4. CD resolution: 0.06 mS/digit and ± 1 digit of measured value.
5. Temperature: 0.05 °C/digit and ± 1 digit of measured value.
Prerequisites for starting and running the test
Error message Description
Power failure Power failure while the test is in progress
Dialines not conn The dialysate lines are not in the interlock shunt.
Shunt Cover open The interlock shunt is open.
Connect Conc.Line
Wrong conc. supply The concentrate connector is in the rinse chamber, or concen-
trate is not connected at all. The error message depends on the
central delivery system preselected in the setup menu.
Blood Sensed by OD The optical detector senses blood in the system.
Flow alarm Line to or from the dialyzer kinked, malfunctions in the hydraul-
ics.
Water alarm Water supply interrupted.
Overview of the individual test steps
Bypass test ..................................................................................................................... 1-8
Optical detector test ........................................................................................................ 1-10
Blood system test............................................................................................................ 1-12
Venous pressure system test ......................................................................................... 1-16
Air detector test............................................................................................................... 1-18
Display test ..................................................................................................................... 1-22
Arterial pressure system test .......................................................................................... 1-24
Battery test ...................................................................................................................... 1-26
Blood leak test ................................................................................................................ 1-28
Temperature test............................................................................................................. 1-30
Negative pressure holding test ....................................................................................... 1-32
Positive pressure holding test ......................................................................................... 1-34
UF function test............................................................................................................... 1-39
Conductivity test ............................................................................................................. 1-42
Diasafe/HDF filter test..................................................................................................... 1-44
Fresenius Medical Care 4008 4/09.03 (TM) 1-7

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 E and is the answer not in the manual?

Fresenius Medical Care 4008 E Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 E
CategoryMedical Equipment
LanguageEnglish

Related product manuals