1-92 Fresenius Medical Care 4008 4/08.03 (TM)
End of hemodialysis
Rinse
Hot rinse
Disinfection (cleaning)
Hemodialysis
PGM 1: –R–
PGM 2: –R– endless
PGM 1: –F–HR–C–
PGM 2: –F–HR–
PGM 3: –IHR– / –IHR–C–
PGM 1: –F–D–M–
PGM 2: –F–HDIS–M–
PGM 3: –F–D–M–HR–
PGM 4: –F–HDIS–M–HR
PGM 5: –F–D(F)–M–
1.3.4 Program runs during the cleaning programs
Fig.: Flow chart of cleaning programs – overview
● Explanation of the abbreviations used
PGM Program
R Rinse
R endless Endless rinse
F Rinsing clear
HR Hot rinsing
C Cooling rinse
D Disinfection
D(F) Disinfection
Disinfectant drawn in from the front (concentrate suction tube).
HDIS Hot disinfection
M Mandatory rinse
IHR Integrated hot rinsing
● Notes on program runs
At the end of the set program, the rinse chamber is evacuated for approx. 1 min.
Any statements on time refer to the factory setting. Shorter or longer program times can be set at
any time by means of the SET UP menu (see Technical Manual, Chapter 6).