5-20 Fresenius Medical Care 4008 4/08.03 (TM)
䊲
䊲
䊱䊲
CPU2: WR ANALOG OUTP
Confirm
key
O: CPU2_TEMP_DET_ADJ
Detuning, temperature display
IC 11/2, 0 to 10 V (X632/A23, 0 to
10 V)
䊱䊲
O: CPU2_DIAL_DET_ADJ
Detuning, dialysate pressure display
IC 11/1, 0 to 10 V (X632/C 20, 0 to
10 V)
䊱䊲
O: CPU2_P_ADS_DET
Not used
IC 11/20, 0 to 10 V (X632/A 20, 0 to
10 V)
䊱䊲
O: CPU2_PV_DET
Detuning, venous pressure
IC 11/20, 0 to 10 V (X632/C 18, 1 to
9V)
䊱䊲
O: CPU2_PA_DET
Detuning, arterial pressure
IC 11/19, 0 to 10 V (X632/A 17, 4 to
7V)
䊱䊲
O: CPU2_COND_DET
Detuning, CD display
IC 11/19, 0 to 10 V (X632/A 21, 0 to
10 V)
O: CPU2_HIGH_RES_OP
OP control voltage
for higher resolution of dialysate pres-
sure
IC 11/20, 0 to 10 V
䊱䊲
䊲
䊲
back to menu ?
Confirm
key
CPU2: WR ANALOG OUTP
5.8 Writing the analog outputs of CPU II
Explanation:
UF Rate display: DAC value (can be changed with: 4008/E/B: ▲▼ UF Rate, 4008 H/S: +/–)
Time Left display: Analog voltage on P.C.B. LP 632, in 0.1 V

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 E and is the answer not in the manual?

Fresenius Medical Care 4008 E Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 E
CategoryMedical Equipment
LanguageEnglish

Related product manuals