EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>4008 E

Fresenius Medical Care 4008 E User Manual

Fresenius Medical Care 4008 E
595 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #150 background imageLoading...
Page #150 background image
Fresenius Medical Care 4008 4/08.03 (TM) 3-15
Ultrafiltration
Reset
Volume
UF
Prog.I/O
UF Volume ml
UF Rate ml/h
UF Goal ml
Time Left h:min
0119
0047
0204
0236
3.2 Calibration mode
3.2.1 Basic conditions
– The hemodialysis machine must be
switched off.
– The service switch must be in the ON
(up) position.
– Turn the hemodialysis machine on.
3.2.2 Messages on the displays on the UF monitor (4008 E/B) or on the screen (4008 H/S)
The values indicated on the display must
be divided by the factor 10.
Values in brackets: tolerances
Watchdog supply voltage
(4.5 V to 5.5 V)
12-V voltage (11.5 V to 12.5 V)
24-V voltage (23 V to 25 V)
Battery charging voltage (>20 V)
7890
4
1
56
23
C Best
Esc
CALIBRATION
004.7 Volt
011.9 Volt
023.6 Volt
020.4 Volt
12 V voltage (11.5 V – 12.5 V)
Watchdog supply voltage (4.5 V – 5.5 V)
Battery charging voltage (> 20 V)
24 V voltage (23 V – 25 V)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 E and is the answer not in the manual?

Fresenius Medical Care 4008 E Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 E
CategoryMedical Equipment
LanguageEnglish

Related product manuals