2-30 Fresenius Medical Care 4008 4/09.03 (TM)
No. Description Measured ✓
value
8 Checking the electrical safety
In Germany according to DIN VDE 0751 standard –1
st
Edition 10/2001.
In other countries, observe the local regulations!
8.1 Visual inspection performed according to item 1. ❏
8.2 Protective earth resistance maximum 0.3 ohms (with power cord) ………… Ω ❏
8.3 Measurement of the leakage current (device leakage current) ❏
❏ Differential current measurement according to fig. C.6
or
❏ Direct measurement according to fig. C.5
Nominal voltage of power supply: ………… V
Device leakage current mains polarity 1 ………… µA
for line voltage ………… V
scaled to nominal voltage (maximum 500 µA, see also Additional conditions) ………… µA
Device leakage current mains polarity 2 ………… µA
for line voltage ………… V
scaled to nominal voltage (maximum 500 µA, see also Additional conditions) ………… µA
Test equipment used ……………………………………………………………
9 Functional test
9.1 T1 test performed – ❏
TSC checklist
for the Technical Safety Checks
to be performed annually (every 12 months)
Page 2/2
4008
Remarks:
Date: Signature: Stamp:
The system has been released for further use ❏ Yes ❏ No
Date: Signature: Stamp: