2-30 Fresenius Medical Care 4008 4/09.03 (TM)
No. Description Measured
value
8 Checking the electrical safety
In Germany according to DIN VDE 0751 standard –1
st
Edition 10/2001.
In other countries, observe the local regulations!
8.1 Visual inspection performed according to item 1.
8.2 Protective earth resistance maximum 0.3 ohms (with power cord) …………
8.3 Measurement of the leakage current (device leakage current)
Differential current measurement according to fig. C.6
or
Direct measurement according to fig. C.5
Nominal voltage of power supply: ………… V
Device leakage current mains polarity 1 ………… µA
for line voltage ………… V
scaled to nominal voltage (maximum 500 µA, see also Additional conditions) ………… µA
Device leakage current mains polarity 2 ………… µA
for line voltage ………… V
scaled to nominal voltage (maximum 500 µA, see also Additional conditions) ………… µA
Test equipment used ……………………………………………………………
9 Functional test
9.1 T1 test performed
TSC checklist
for the Technical Safety Checks
to be performed annually (every 12 months)
Page 2/2
4008
CareFreseniusMedical
Remarks:
Date: Signature: Stamp:
The system has been released for further use Yes No
Date: Signature: Stamp:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 4008 E and is the answer not in the manual?

Fresenius Medical Care 4008 E Specifications

General IconGeneral
BrandFresenius Medical Care
Model4008 E
CategoryMedical Equipment
LanguageEnglish

Related product manuals