6-10 Fresenius Medical Care 4008 4/08.03 (TM)
● Part 1: Set alarm and warning time
SET ALARM/WARN TIME
Confirm
key
Set ART-AL DELAYTIME
Confirm
key
see Part 1.
Set VEN-AL DELAYTIME
Confirm
key
see Part 1.
Set FLOW-OFF W-TIME
Confirm
key
see Part 1.
Set UF-WARNING-TIME
Confirm
key
see Part 1.
Set MUTE-TIME
Confirm
key
see Part 1.
back to menu ?
Confirm
key
● Part 1.1: Set the delay time of the arterial alarm
Set ART-AL DELAYTIME
Confirm
key
Art Al Delay = 5s
Override
key
DATA STORED
After approx. 3 sec
Select
key
Set the desired delay time
(0 to 5 sec) by pressing
the ▲▼ keys.
● Part 1.2: Set the delay time of the venous alarm
Set VEN-AL DELAYTIME
Confirm
key
Ven Al Delay = 5s
Override
key
DATA STORED
After approx. 3 sec
Select
key
Set the desired delay time
(0 to 5 sec) by pressing
the ▲▼ keys.