6-10 Fresenius Medical Care 4008 4/08.03 (TM)
Part 1: Set alarm and warning time
▲▼
▲▼
▲▼
▲▼
▲▼
▲▼
SET ALARM/WARN TIME
Confirm
key
Set ART-AL DELAYTIME
Confirm
key
see Part 1.
1
Set VEN-AL DELAYTIME
Confirm
key
see Part 1.
2
Set FLOW-OFF W-TIME
Confirm
key
see Part 1.
3
Set UF-WARNING-TIME
Confirm
key
see Part 1.
4
Set MUTE-TIME
Confirm
key
see Part 1.
5
back to menu ?
Confirm
key
Part 1.1: Set the delay time of the arterial alarm
Set ART-AL DELAYTIME
Confirm
key
Art Al Delay = 5s
Override
key
DATA STORED
After approx. 3 sec
Select
key
Set the desired delay time
(0 to 5 sec) by pressing
the ▲▼ keys.
Part 1.2: Set the delay time of the venous alarm
Set VEN-AL DELAYTIME
Confirm
key
Ven Al Delay = 5s
Override
key
DATA STORED
After approx. 3 sec
Select
key
Set the desired delay time
(0 to 5 sec) by pressing
the ▲▼ keys.