BATTERY CARE AND MAINTENANCE
K0553 Revision D 122 [EN] English
Battery care and maintenance
ADTS Touch battery pack
The battery unit contains a Lithium Ion battery which requires no user maintenance.
Prolonged exposure to temperature extremes may significantly reduce battery lifetime. For
maximum lifetime, avoid prolonged periods where the battery is exposed to temperatures
outside the range -30°C to +45°C.
The recommended storage temperature range is 5°C to 21°C (41°F to 98.8°F).
The battery is removable. The manufacturer makes the following safety recommendations. The
User should not:
Short circuit the battery.
Immerse the battery in any liquid.
Disassemble or deform the battery.
Expose to, or dispose of the battery in a fire.
Subject the battery to excessive physical shock or vibration outside the specified limits for
the ADTS.
Use a battery that appears to have suffered abuse.
The battery unit is sealed for life, so leakage of electrolyte is not expected. If leakage of electro-
lyte from the battery is observed, stop using the battery immediately and avoid contact with the
electrolyte. If skin or clothes come into contact with the electrolyte, wash immediately with soap
and water. If electrolyte comes into contact with the eyes, wash the eyes thoroughly with water
and consult a Doctor immediately.

Table of Contents

Other manuals for GE ADTS552F

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE ADTS552F and is the answer not in the manual?

GE ADTS552F Specifications

General IconGeneral
BrandGE
ModelADTS552F
CategoryTest Equipment
LanguageEnglish

Related product manuals