2.3 Emergency procedures
In this section
This section describes how to do an emergency shutdown of the ÄKTAprime plus system.
The section also describes the result in the event of power failure.
Emergency shutdown
In an emergency situation, do as follows to stop the run:
ActionStep
To pause the run without ending the method, press the Pause button located
at the instrument front.
1
If required, switch off power to the instrument by pressing the Main power
switch to the 0 position. The run is interrupted immediately.
2
Power failure
The result of a power failure depends on which unit that is affected.
will result in...Power failure to...
The run is interrupted immediately, in an undefined
state
The data collected up to the time of the power failure
is available in PrimeView
ÄKTAprime plus
The PrimeView computer shuts down in an undefined
state
The run continues, but data cannot be saved in
PrimeView.
Computer
26 ÄKTAprime plus Operating Instructions 28-9597-89 AB
2 Safety instructions
2.3 Emergency procedures

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AKTAprime plus and is the answer not in the manual?

GE AKTAprime plus Specifications

General IconGeneral
BrandGE
ModelAKTAprime plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals