GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-II
X-RAY
UNIT
7-1576
SM
A0654E
SEC.
17
MOTOR
DRIVE
SYSTEM
initial
Conditions:
@
Main
power
circuit
breaker
-
ON
©
X-ray
generator
keyed
switch
-
OFF
SPEED
DRIVE
e
=
Charger
cord
disconnected.
CONTROL
BRAKE
ASSIST
LATCH
KNOB
HANDLE
BUTTON
BAR
17-1
Motor
Drive
Controls
and
Safety
Interlocks
1.
Check
Speed
Control
Knob.
The
Drive
Motor
Speed
Control
Knob
should
rotate
freely
and
spring
return.
2.
Place
Direction
Selector
in
FORWARD
position.
Depress
Brake
Handle
and
advance
Speed
Con-
trol
Knob.
The
slow
speed
should
begin
with
some
pre-travel
of
the
speed
lever.
3.
Check
Brake
Handle.
The
brake
handle
limit
switch
should
be
actuated
when
the
Brake
Handle
is
depressed
within
3/8”
(9.53
mm)
of
its
bottom,
brake
released,
position.
4.
Check
front
safety
bumper.
The
bumper
should
depress
completely
without
binding
and
spring
return.
The
drive
power
should
be
de-energized
when
the
bumper
is
depressed
approximately
1/4
inch.
(6.35
mm).
5.
Plug
the
changer
cord
into
a
convenience
outlet.
Check
that
the
motor
drive
system
is
disabled
when
the
charger
cord
is
plugged
into
a
conven-
ience
outlet.
23573-P10
DIRECTION
CHARGER
SAFETY
6.
Turn
circuit
breaker
-
OFF.
SELECTOR
CORD
BUMPER
ILLUSTRATION
4-11-51
4-11-41

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE AMX-II and is the answer not in the manual?

GE AMX-II Specifications

General IconGeneral
BrandGE
ModelAMX-II
CategoryMedical Equipment
LanguageEnglish

Related product manuals