GENERAL
ELECTRIC
MEDICAL
SYSTEMS
AMX-ll
X-RAY
UNIT
1-1880
SM
A0654E
SECTION
19
BATTERY
CAPACITY
CHECK
initial
Conditions:
Main.
power
circuit
breaker
-
ON
@
X-ray
generator
keyed
switch
-
ON
e
Battery
charged
and
allowed
to
rest
15
minutes
before
this
test
instrumentation:
@
Battery
voltage:
digital
voltmeter
connected
across
3TB1-27
(+)
and
3TB1-5
(-).
1.
Select:
100
KVP,
100
MAS.
Close
collimator
blades
and
shield
with
lead.
4.4
Take
one
exposure
every
minute
for
10
minutes. Do
not
run
rotor
more
than
6
seconds
per
exposure.
Record
battery
voltage
after
each
exposure.
After
the
series
of
10
exposures,
allow
batteries
to
rest
for
one
minute
with
no
power
being
drawn.
If
battery
voltage
is
less
than
112
VDC
after
one
minute,
battery
should
be
reconditioned
as
des-
cribed
in
Chapter
8,
Section
5-4.
Pe
5.
Turn
keyed
switch
and
circuit
breaker
OFF.
Disconnect
digital
voltmeter,
and
secure
close
operator’s
panel.
Repiace
x-ray
control
cover,
installing
6-screws
holding
it.
Replace
front
and
rear
covers.
411-45