Troubleshooting: Monitor defaults transfer
2000966-542D Dash 3000/4000/5000 6-11
Monitor defaults transfer
Below is a list of error messages that may display when performing a
patient monitor defaults transfer.
Storing monitor defaults
Copying stored monitor defaults
Message Description
ERROR - UNIT DEFAULTS HAVE NOT
BEEN SAVED
Patient monitor defaults were not saved on
the server patient monitor due to a memory
error in the patient monitor.
Message Description
ERROR COPYING UNIT DEFAULTS -
NETWORK ERROR
Either an error has occurred while the
defaults were being transferred, or defaults
have not been saved on the server patient
monitor.
ERROR COPYING UNIT DEFAULTS -
UNIT TYPE MISMATCH
The server and client patient monitors are
set to different unit types (e.g., ADULT,
NEO, or OR)
ERROR COPYING UNIT DEFAULTS -
SOFTWARE VERSION MISMATCH
The server and client patient monitors have
different software versions (e.g., V2C, V3A).
ERROR COPYING UNIT DEFAULTS -
DATA CORRUPTION ERROR
The patient monitor defaults transferred have
become corrupted, possibly due to a
memory error.
ERROR COPYING UNIT DEFAULTS -
SOFTWARE FEATURE LEVEL
MISMATCH
The server and client patient monitors are
set to different software levels (e.g., Basic,
Cardiac, Cardiopulmonary).
ERROR COPYING UNIT DEFAULTS -
COUNTRY CODE MISMATCH
The server and client patient monitors are
set to different country codes (e.g.,
DEFAULT or FRANCE).

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
BrandGE
ModelDash 4000
CategoryMonitor
LanguageEnglish

Related product manuals