Functional Checks
C
O
N
F
I
N
S
T
C
M
P
M
P
A
S
S
F
A
I
L
Monitor Power Up - (POST)
ECG 5 LD (S)
ECG 12SL
(option)
Respiration (S)
Temperature 1 and 2 (S)
Cardiac Output
(option)
Invasive BP
(option)
Pulse Oximetry (S)
NBP (S)
Analog Out, Defibrillator
Synchronization
(S)
End-Tidal CO2
(option)
Batteries
Battery A
(option)
Battery B
(option)
Printer
(option)
Display (S)
Speaker (S)
Network
(option)
Remote Control
(option)
BISx
(option)
Wireless LAN
(option)
Dash Port 2 Docking
Station
(option)
TRAM-RAC 2A w/ICG
(option)
TRAM-RAC 2A w/SAM
(option)
Abbreviations
DUT Device Under Test
Standard (S) Standard configuration test required
Option (O) Test as required if the option is installed
INST Installation
CM Corrective Maintenance
PM Preventive Maintenance
CONF Configuration – Standard (S) or Option (O)
Device Test Results Tested By Date
Pass (9) Fail (9)
Step
DUT
Done
9
Electrical Safety Tests
Test
Results
Expected
Limits
I
N
S
T
C
M
P
M
P
A
S
S
F
A
I
L
A Ground Continuity Ohms Ohms
1. Measured Resistance
<.2
B Earth Leakage Current
µ
A
µ
A
USA
1. Pol Norm, Neut Clsd
300 500
2. Pol Norm, Neut Open
300 1000
3.
Pol Rev, Neut Open
300 1000
4.
Pol Rev, Neut Clsd
300 500
C Enclosure Leakage Current
1. Pol Norm, Neut Clsd, Gnd Clsd
100
2. Pol Norm, Neut Open, Gnd Clsd
300
3. Pol Norm, Neut Clsd, Gnd Open
300
4. Pol Rev, Neut Clsd, Gnd Open
300
5. Pol Rev, Neut Open, Gnd Clsd
300
6. Pol Rev, Neut Clsd, Gnd Clsd
100
D
Patient Leakage Current
to Ground
ECG
SP
O2
µ
A
1. Pol Norm, Neut Clsd, Gnd Clsd
10
2. Pol Norm, Neut Open, Gnd Clsd
50
3. Pol Norm, Neut Clsd, Gnd Open
50
4. Pol Rev, Neut Clsd, Gnd Open
50
5. Pol Rev, Neut Open, Gnd Clsd
50
6. Pol Rev, Neut Clsd, Gnd Clsd 10
E.
Patient Leakage Current
Mains on Applied Part
ECG
SP
O2
µ
A
1. Pol Norm, Neut Clsd, Gnd Clsd
50
2. Pol Rev, Neut Clsd, Gnd Clsd
50
F.
BISx Patient Leakage Current
to Ground (option)
µ
A
µ
A
1. Pol Norm, Neut Clsd, Gnd Clsd
100
2. Pol Norm, Neut Open, Gnd Clsd
500
3. Pol Norm, Neut Clsd, Gnd Open
500
4. Pol Rev, Neut Clsd, Gnd Open
500
5. Pol Rev, Neut Open, Gnd Clsd
500
6. Pol Rev, Neut Clsd, Gnd Clsd
100
G.
BISx Patient Leakage Current
Mains on Applied Part (option)
µ
A
µ
A
1. Pol Norm, Neut Clsd, Gnd Clsd
5000
2. Pol Rev, Neut Clsd, Gnd Clsd
5000