Troubleshooting: Review errors
6-14 Dash 3000/4000/5000 2000966-542D
Useful error data
Below is sample error log followed by a description of parameters found
in the error log.
Process Name
The name of the software task that was operating when the event/
problem occurred.
Error Code
The error log contains more than just operating system errors. Many
events that have an impact upon the system are also entered into the log.
The 700-series of error codes are system-initiated events. Listed below
are some of the event/error codes you might find useful.
049A
This information is only
useful to the manufacturer.
Definition of error codes
Error code Description
400-4FF Network errors were detected.
703 Diagnostic tests were completed.
70B Internet address was changed. The network address for the patient
monitor was changed. This should only be done by qualified service
personnel.
70C Video test was completed. This test should only be performed by
qualified service personnel.
70E Time was changed from this patient monitor. This helps determine
how the system-wide time may have been altered.

Table of Contents

Other manuals for GE Dash 4000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Dash 4000 and is the answer not in the manual?

GE Dash 4000 Specifications

General IconGeneral
BrandGE
ModelDash 4000
CategoryMonitor
LanguageEnglish

Related product manuals