GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 2 5176463-100
19
7. Keep a record of the results with other hand copies of PM data kept on site.
2-3-3 Chassis Leakage Current Test (Continued)
Date sheet for Chassis Source Leakage Current
Follow the foregoing test procedure. The test passes when all readings measure less than
the value shown in Table 2-4 below.
Table 2-3
MAXIMUM ALLOWANCE LIMIT FOR CHASSIS SOURCE LEAKAGE
CURRENT
Country Normal
Condition
Open Ground Reverse
Polarity
Open Neutral
USA N/A 0.3mA 0.3mA N/A
Others 0.1mA 0.5mA 0.5mA 0.5mA
Caution:
When the meter’s ground switch is OPEN, do not touch the unit to avoid
possible electric shock.
Table 2-4
DATA SHEET FOR CHASSIS SOURCE LEAKAGE CURRENT
Unit
Power
Tester
POLARIT
Y Switch
Tester
NEUTRA
L Switch
Test 1 Probe
Connector
Test 2 Real
Panel
Test 3 Mounting screw
on the handle
ON NORM OPEN
ON NORM CLOSED
ON REV OPEN
ON REV CLOSED
OFF NORM OPEN
OFF NORM CLOSED
OFF REV OPEN
OFF REV CLOSED