GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 2 5176463-100
36
6-4-1 Removal of the retainer ring of trackball
1. Turn off power supplyand pull out power plug of system
2. Infix one end of convexity substance of trackball to the guiding hole of retainer ring
(2 holes).
3. Turn the trackball according to counter-clockwise slowly.
4. Take retainer ring out of keyboard.
5. Take out the trackball.
6-4-2 Cleaning the trackball
Wipe any oil or dust on the trackball with TFC dry cleaner or dry cloth.
Note:
Keep retainer ring and trackball in certain place to avoid losing.
Note:
When clean trackball tee, do not let any liquid into it (system/keyboard). Clean
the trackball parts with alcoholpropyl alcohol. Don’t use anyother impregnant to
avoid damaging rubber and mechanical parts of trackball.
6-4-3 Fixing the trackball and retainer ring
1. Put in the trackball.
2. Put in the retainer ring.
3. Aim trackball to the guiding hole, then fix retainer ring according to clockwise.
Note:
Don’t use huge strength to fix the retainer ring to avoid damage.
Note:
If there is need for maintenance tool or information of trackball, turn to sales’
person or service person for help.
1. It is suggested to use 2% diacid aldehyde liquid without surface activation (for
example Cidex.
2. Do not use alcohol or any liquid like alcohol.
3. Wipe off any leftover in the probe with distilled water after antisepsising with
antiseptic.