GE MEDICAL SYSTEMS LOGIQ A1 BASIC SERVICE MANUAL
REV 2 5176463-100
35
5. Do not use liquid olefin, petroleum generic coupling liquid or other unapproved
materialsas these materials can damage the probe.
6. Do not put probe into any proportion immerge or dunk in the any material’s
cleanser. Do not use autoclaveultraviolet radiationgas or heat treatment
antisepsis technics to antisepsising probe.
7. When probe is immerged appropriate industry antiseptic, use for clinic medicine
applied registered EPA antiseptic. According to user operation’s note proceeding
chromacontiguity timestockpile and deal with operation of antisepsising
manufacture.
8. It is suggested to use 2% diacid aldehyde liquid without surface activation (for
example Cidex.
9. Do not use alcohol or any liquid like alcohol.
10. After use antiseptic to disinfect, use distilled water to wash all hangover in the
probe.
6-3-3 Testing
1. Make sure that the hospital’s power supply is regular.
2. Insure all power voltage is within the prescribe range. See chapter 4, function of
maintain adjust procedure examine.
3. Carry out electric safety testing, the contents are as follows:
• Hospital power socketTest of fault of power connection
• All commutatorsTest of commutator creep age
• Mainframe and external device (Test of creep age in shell and grounded anti-
resistance)
6-4 Trackball maintenance
Keep the trackball clean to make it work well. If there is dust in, it might work
abnormally or might be damaged.
Tools of trackball maintenance:
1. Trackball tooltools for opening
2.
Cotton stickeralcoholpropyl-alcohol or dry cloth.