GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
10-10 Section 10-5 - System Maintenance
10-5-8 Battery Performance Maintenance
Battery replacement every three years is recommended.
It is recommended to do battery performance maintenance one time per year.
Please follow the flow chart below to carry out battery performance maintenance.
NOTE: Disconnect all probes when discharge battery.
NOTE: Discharge the battery to let the system automatically shut down.
Figure 10-1 Flow chart of Battery Performance Maintenance
Battery discharge
Battery charge > 3 hour
Yes
No
Battery power > 90%?
Battery charge > 3 hour
Battery power > 90%?
No
Battery is Ok
Yes
Yes
Discharge Replace battery
No
Yes
Battery is Ok
Replace battery
>45 minutes?
Replace battery
>45 minutes?
No