General procedures
LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-3
5726264-100 English Rev.8
4-2 General procedures
Ultrasound system requires all covers.
Operate this Ultrasound system only when all board covers and
frame panels are securely in place. The covers are required for
safe operation, good Ultrasound system performance and
cooling purposes.
Energy Control and Power Lockout for LOGIQ V3/V5/V5
Expert.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.