General Ultrasound system requirements
LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-3
5726264-100 English Rev.8
2-2 General Ultrasound system
requirements
2-2-1 Contents in this section
• 2-2-2 ‘Ultrasound system environmental requirements’ on
page 2-3
• 2-2-3 ‘Electrical requirements’ on page 2-6
• 2-2-4 ‘EMI limitations’ on page 2-9
• 2-2-5 ‘Probes environmental requirements’ on page 2-11
• 2-2-6 ‘Time and manpower requirements’ on page 2-11
2-2-2 Ultrasound system environmental requirements
2-2-2-1 If the Ultrasound system is very cold or hot
When unpacking the Ultrasound system, allow the temperature
of the Ultrasound system to stabilize before powering up. The
following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
CAUTION
If the Ultrasound system is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Table 2-1: System Acclimation Time Chart
Degree C -4.5 -2 0.5 3 40 42.5 45 47.5 50 55 60
Degree F 23.9 28.4 32.9 37.4 104 108.5 113 117.5 122 131 140
hours
3
2
1
0 0 1 23 468

Table of Contents

Other manuals for GE LOGIQ V3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V3 and is the answer not in the manual?

GE LOGIQ V3 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V3
CategoryMedical Equipment
LanguageEnglish

Related product manuals