Global Service User Interface (GSUI)
LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-15
5726264-100 English Rev.8
7-4-2 Control Frame
Contains the user interface elements used for:
• Diagnostic control, and
• Operator feedback
7-4-2-1 Button
This button has two modes each with appropriate text:
Execute - to start the diagnostic, and
Abort - to stop a diagnostic
The button can also be disabled.
7-4-2-2 Loop Count
This is an editable text field that will only accept numeric values
with 4 digits or less. When the button is configured as an
“execute” button and pressed, the loop count field will be
queried to determine the number of times to execute the
diagnostic.
7-4-2-3 Progress Indicator
Displays a graphical progress indication to the user.
7-4-2-4 Short Text Message
Displays either a starting message or aborting message, as well
as the diagnostic completion status.
7-4-2-5 Background Color
Initially gray, the Control Frame background color changes upon
completion of a diagnostic to indicate completion status.
• Fail = Red
Pass = Green
Neither pass nor fail = Set back to Gray (for example, final
code status is Aborted).

Table of Contents

Other manuals for GE LOGIQ V3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V3 and is the answer not in the manual?

GE LOGIQ V3 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V3
CategoryMedical Equipment
LanguageEnglish

Related product manuals